Description
SPECIFICATIONS :
| Light source | : | 300W Xenon lamp |
| Filter | : | Air mass filter (AM1.5G) |
| Solar Simulator Class | : | A C A |
| Spectral Match | : | 0.75 – 1.25 |
| Spatial non-uniformity of irradiance | : | 10 % |
| Temporal instability of irradiance | : | 2 % |
| Maximum beam diameter | : | 50 mm |
| Maximum intensity | : | 1 Sun |
| Minimum intensity | : | 0.7 Sun |
| Bias voltage | : | +1V to -1V |
| Bias voltage resolution | : | 5 mV |
| Current measurement range | : | 0.1 uA to 0.75 A |
| Sample size | : | Up to 50 mm |
| Sample holder with XY motion | ||
| Dimensions | : | 600 x 922 x 525 mm ( W x H x D) |
Xenon Lamp Specifications :
| Quartz Type | : | Ozone Free |
| Wattage | : | 300 W |
| Luminous Flux | : | 7000 lm |
| Luminous Intensity | : | 700 cd |
| Average Luminance | : | 27000 cd / cm2 |
| Regulation Mode | : | Constant current with power limit |
| Protection | : | Open / Short circuit proof, overtemp, over current |
Software Features :
- Performs IV scanning in user specified range
- Exports the results to spreadsheet file
- Operating system requirement: Windows XP / 7 / 8 / 8.1 / 10
- PC Interface: RS232 or USB (Using a USB to serial converter)
Pour en savoir plus sur les produits Holmarc, veuillez cliquer ici